Rathinam, Sridhar and Grobler, Sophia and Bleetman, Antony and Kink, Thomas and Steyn, Richard (2014) Evolved design makes ThoraQuik safe and user friendly in the management of pneumothorax and pleural effusion. Emergency medicine journal : EMJ, 31 (1). pp. 59-64. ISSN 1472-0213. This article is accessible to all HEFT staff and students via NHS Evidence www.evidence.nhs.uk by using their HEFT Athens login IDs

Full text not available from this repository.
Official URL: http://emj.bmj.com/content/31/1.toc

Abstract

BACKGROUND

We have previously described the utility of ThoraQuik, a device designed to be fit for purpose for aspirations of pneumothorax and pleural effusions. We evaluated the safety, efficacy and operator handling of the evolved prototype, ThoraQuik II, which has a lesser profile and a spring loaded Veres needle for added safety.

METHODS

A prospective, observational clinical trial with ethics and MHRA approval was conducted in a single centre. Patients with diagnosed pneumothorax (including tension pneumothorax) and pleural effusion were consented and recruited. The ease of device introduction, penetration and ease of use were evaluated. Clinical and radiological improvements were the clinical endpoints and operator feedback was analysed.

RESULTS

20 procedures were performed on patients (mean age: 63.4 years (range: 30-90 years) with 75% male subjects) recruited between September 2008 and August 2009. Nine patients had pneumothorax (tension pneumothorax n=4) and 11 had pleural effusions. 19 patients completed the study with symptomatic and radiological resolution. One patient was withdrawn due to poor pain threshold disproportionate to the procedure. No complications were encountered. 68% had complete clinical and radiological resolution and 32% had partial resolution (these patients needed a definitive drain and hence were not aspirated to completion). The operator feedback in the study rated the device as very good or excellent in 90% patients.

CONCLUSIONS

Our study found the use of ThoraQuik II to be safe and easy in draining pneumothorax and pleural effusions. The changes to ThoraQuik II made it more user friendly.

Item Type: Article
Additional Information: This article is accessible to all HEFT staff and students via NHS Evidence www.evidence.nhs.uk by using their HEFT Athens login IDs
Subjects: WG Cardiovascular system. Cardiology
Divisions: Planned IP Care > Thoracic Surgery
Related URLs:
Depositing User: Mrs Yolande Brookes
Date Deposited: 03 Jul 2014 14:07
Last Modified: 03 Jul 2014 14:07
URI: http://www.repository.heartofengland.nhs.uk/id/eprint/329

Actions (login required)

View Item View Item