Long-Term Follow-Up of Isolated Epicardial Left Ventricular Lead Implant Using a Minithoracotomy Approach for Cardiac Resynchronization Therapy.

McAloon, Christopher J and Anderson, Benjamin M and Dimitri, Wadih and Panting, Jonathan and Yusuf, Shamil and Bhudia, Sunil K and Osman, Faizel (2016) Long-Term Follow-Up of Isolated Epicardial Left Ventricular Lead Implant Using a Minithoracotomy Approach for Cardiac Resynchronization Therapy. Pacing and clinical electrophysiology : PACE, 39 (10). pp. 1052-1060. ISSN 1540-8159. This article is available to all HEFT staff and students via http://tinyurl.com/z795c8c by using their HEFT Athens login IDs

Full text not available from this repository.
Official URL: http://onlinelibrary.wiley.com/doi/10.1111/pace.12...

Abstract

BACKGROUND

Transvenous left ventricular (LV) lead placement for cardiac resynchronization therapy is unsuccessful in 5-10% of reported cases. These patients may benefit from isolated surgical placement of an epicardial LV lead via minithoracotomy approach.

AIM

To evaluate the success of this approach at long-term follow-up.

METHODS

Retrospective evaluation of all consecutive patients undergoing isolated epicardial LV lead placement after failed transvenous attempt over a 6-year period. Data collected on baseline parameters, procedural details, and outcome at follow-up (hospital stay, complications, mortality, and clinical response).

RESULTS

Forty-two patients underwent epicardial lead implant. Five died within 1 year (11.9%): two (4.8%) died within 30-days post op (one from intraoperative hemorrhage, the other from multiple organ failure); 39 (95.1%) were admitted to the high dependency unit and transferred to the ward <24 hours. Median hospital stay was 3.4 ± 1.9 days. The overall complication rate was 17.5% (n = 7): 15.0% (n = 6) short term and 2.5% (n = 1) long term; these included three (7.5%) LV noncapture events all treated with reprogramming. There were two (5.0%) wound infections requiring oral antibiotics and two (5.0%) device infections requiring intravenous antibiotics (one had device resiting, the other developed septic shock requiring intensive care admission). Assessment of clinical response was possible in 34 (81.0%) at follow-up: 21 (61.8%) were responders and 13 (28.2%) nonresponders with no significant differences between these groups; no clinical predictors of response were identified.

CONCLUSION

Isolated epicardial LV lead implant using minithoracotomy is relatively safe and effective at successful LV pacing. Response rate and postoperative recovery at long-term follow-up are reasonable in these high-risk patients.

Item Type: Article
Additional Information: This article is available to all HEFT staff and students via http://tinyurl.com/z795c8c by using their HEFT Athens login IDs
Subjects: WG Cardiovascular system. Cardiology
Divisions: Emergency Services > Cardiology
Related URLs:
Depositing User: Mrs Caroline Tranter
Date Deposited: 24 Feb 2017 14:32
Last Modified: 24 Feb 2017 14:32
URI: http://www.repository.heartofengland.nhs.uk/id/eprint/1207

Actions (login required)

View Item View Item